How to use the ECG247 portal
ECG247 portal is intended for healthcare professionals with knowledge of ECG and arrhythmias.
How to use the ECG247 portal
How to interpret an ECG247 test
ECG247 Portal User Guide
ECG Interpretation
Practise interpreting long-term ECGs – and spotting when the automatic algorithm gets it wrong. Choose a case, give your interpretation, and get a reasoned review from JARLE.
Cases
Loading cases … If this text remains, the JavaScript code has not been allowed to run.
Automatic arrhythmia detection
ECG247 detects all common heart rhythm disturbances
Rhythm disturbances are detected and colour-coded automatically, so the most urgent findings are always shown first.
Always verify. All automatically detected rhythm disturbances must be verified by a doctor before starting treatment.
- Ventricular tachycardia ≥30 s
- Pause ≥4 s
- Heart rate <30/min (≥30 s)
- Supraventricular tachycardia ≥30 s
- Atrial fibrillation ≥30 s
- Atrial flutter ≥30 s
- Ventricular tachycardia ≥4 beats and <30 s
- Supraventricular tachycardia 15–30 s
- Heart rate >180/min (≥5 min)
- Ventricular ectopic beats
- Supraventricular ectopic beats
- Sinus arrhythmia / irregular beats
- Heart rate >180/min (<5 min)
- Periodic recordings
Recording length
ECG files are recorded in standard one-minute intervals. During ongoing arrhythmias the recording is extended to four minutes, and continuous arrhythmias are stored in successive four-minute files.
Manual recordings – coded black
Recordings started manually are coded black. If an arrhythmia is detected, the recording is shown in the relevant arrhythmia category. Manual recordings should always be interpreted.